Evaluation of a capillary gas chromatographic impurity test procedure for 4-hexylaniline
1University of Cincinnati, Mail Location 0172, Cincinnati, OH 45221-0172, USA. bhymerc@email.uc.edu
Abstract:
A capillary gas chromatographic test procedure for the detection and quantitation of impurities in the bulk intermediate, 4-hexylaniline, is evaluated and found to be accurate and precise. 4-Hexylaniline is dissolved in methanol and chromatographed isothermally at a temperature of 195 degrees C on a 60-m x 0.32-mm 85% polyethylene glycol-15% dimethylsilicone blend (DX-4) film column. A flame ionization detector is used, and the impurities in the parent compound are estimated from peak areas on a percent basis compared with the area of the parent peak in the chromatogram. Response factors are determined for the known impurities. Validation of this test method includes a recovery study of known impurity spiked samples fortified in the range of 0.1-1% (w/w). A repeatability study is performed, consisting of the analysis of two different synthetic batch lots of 4-hexylaniline analyzed over three experimental run days using two chromatographic columns of different manufacturing lots. These data and other aspects of this test procedure are discussed.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
High-Performance Liquid Chromatography: Elution Process