Related Experiment Video
Updated: Aug 17, 2026

Performing Behavioral Tasks in Subjects with Intracranial Electrodes
Published on: October 2, 2014
The potential value of three-dimensional accelerometry for detection of motor seizures in severe epilepsy
Tamara M E Nijsen1, Johan B A M Arends, Paul A M Griep
1Epilepsy Centre Kempenhaeghe, Sterkselseweg 65, 5591 VE Heeze, The Netherlands. NijsenT@Kempenhaeghe.nl
Abstract:
Seizure detection results based on the visual analysis of three-dimensional (3D) accelerometry (ACM) and video/EEG recordings are reported for 18 patients with severe epilepsy. They were monitored for 36 hours during which 897 seizures were detected. This was seven times higher than the number of seizures reported by nurses during the registration period. The results in this article demonstrate that 3D ACM is a valuable sensing method for seizure detection in this population. Four hundred twenty-eight (48%) seizures were detected by ACM. With 3D ACM alone it was possible to detect all the seizures in 10 of the 18 patients. Three-dimensional ACM also was complementary to EEG in our population. ACM patterns during seizures were stereotypical in 95% of the motor seizures. These characteristic patterns are a starting point for automated seizure detection.
More Related Videos
10:22Interictal High Frequency Oscillations Detected with Simultaneous Magnetoencephalography and Electroencephalography as Biomarker of Pediatric Epilepsy
Published on: December 6, 2016
06:58Non-restraining EEG Radiotelemetry: Epidural and Deep Intracerebral Stereotaxic EEG Electrode Placement
Published on: June 25, 2016