Valve dysfunction of the cloth-covered Starr-Edwards ball valve
Motomi Shiono1, Akira Sezai, Mitsumasa Hata
1Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan. mshiono@med.nihon-u.ac.jp
Background:
Between June 1968 and March 1977, Starr-Edwards cloth-covered ball valves were used for valve replacement on a routine basis.
Methods And Results:
Among the 66 operative survivors who underwent an isolated aortic or mitral valve replacement, 20 patients required reoperation 22 times because of valve dysfunction, thromboembolic complication, paravalvular leakage, hemolytic anemia, and/or prosthetic valve endocarditis. Reoperation was performed at a mean of 15.9+/-9.8 years after initial replacement. Excised valves were examined and reoperation after initial operation was reviewed. Operative mortality was 10.0%. Freedom from reoperation for aortic valve replacement and mitral valve replacement was 56.2% at 34 years and 61.0% at 37 years after initial operation, respectively. Cloth wear or pannus formation were observed in all excised prostheses. Orifice cloth was more markedly worn in mitral valves than in aortic valves, particularly in mitral valves of more than 20 years old. Pannus overgrowth contributed to valve regurgitation in the older valves.
Conclusions:
Early diagnosis of valve dysfunction and reoperation are recommended as soon as symptoms appear.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Stenosis IV: Nursing Management
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Mitral Stenosis I: Introduction
Mitral Valve Prolapse I: Introduction
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...