The patient: a novel source of error in clinical temperature measurement using infrared aural thermometry
Andrew I Heusch1, Peter W McCarthy
1Welsh Institute of Chiropractic, University of Glamorgan, Wales, UK.
Introduction:
The vital signs are an important component of patient assessment. With respect to body temperature; there has been a move away from mercury-in-glass thermometers toward the relatively inexpensive, safer to use infrared tympanic, auricular, or ear thermometer. Although already in widespread use, the reliability of these devices has increasingly been called into question. Few studies, however, have considered that the problem might reside outside the device itself.
Objective:
To investigate the correlation between left ear temperature and right ear temperature using an infrared tympanic thermometer (ITT).
Design:
A prospective, single-blind trail with randomized assignment of the first ear to be assessed.
Setting/Location:
Welsh Institute of Chiropractic (WIOC), University of Glamorgan.
Subjects:
One hundred and thirty two (132) asymptomatic subjects who were opportunistically sampled on entry into the WIOC. None of the subjects had any neurologic deficit, or any known underlying pathologic problems. All were students at the University of Glamorgan (age range, 18-48 years). A smaller sample of the same cohort was used on a separate occasion to compare ITTs from 2 different manufacturers using the same protocol.
Intervention:
Measurement of ear (ear canal/tympanic) temperature bilaterally using an ITT.
Outcome Measures:
Bilateral asymmetry in ear canal temperature. Analysis consisted of nonparametric statistical tests, using Spearman's correlation for comparison and the Wilcoxon matched pairs test for differences.
Results:
In 132 subjects who completed the main study, although a degree of correlation was found, the relationship did not extrapolate through zero. At temperatures more than 0.4 degrees C above or below 36.7 degrees C, there was a significant difference in temperature in the left compared with the right ear. Additionally, the left ear tended to register a lower temperature than the right ear at temperatures below 36.7 degrees C, whereas the left ear tended to read a higher temperature than the right ear above 36.7 degrees C.
Conclusions:
The authors suggest that clinicians consider measuring both ears, and take the greater of the readings to be a more accurate reflection of the patient's core temperature. The difference may perhaps even be related to the person's health. This change in methodology should increase the reliability of the measurement and go some way toward restoring confidence in the use of ITT devices.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...