Related Experiment Video
Updated: Aug 17, 2026

Artificial Lung Device Priming for In Situ Fiber Bundle Surface Grafting
Published on: March 28, 2025
Covalently grafted phospholipid monolayer on silicone catheter surface for reduction in platelet adhesion
Ohm Divyam Krishna1, Kwangmeyung Kim, Youngro Byun
1Department of Materials Science & Engineering, Center for Cell and Macromolecular Therapy, Gwangju Institute of Science & Technology, Republic of Korea.
Abstract:
We report a novel method of surface grafting a polymeric phospholipid system containing an acryloyl end group (1stearoyl-2-[12-(acryloyloxy)-dodecanoyl]-sn-glycero-3-phosphocholine) onto medical grade silicone catheters. The surface of silicone catheters was functionalized in a sequence of steps: plasma polymerization of allyl alcohol on the catheter surface, grafting acryloyl moieties and in situ polymerization of the pre-assembled acryloyl terminated phospholipids on the acryloyl functionalized catheter surface. The surface morphological changes analyzed by scanning electron microscopy (SEM) and atomic force microscopy (AFM), a sharp decrease in water contact angle, and appearance of N1s peak in XPS analysis indicated a successful monolayer grafting of the phospholipid. In platelet adhesion tests performed using platelets isolated from rabbit plasma, the phospholipid grafted surface showed fewer adhered platelets, without emerging pseudopodes or aggregation. However, ungrafted catheter surface showed large number of platelets in extensively spread and aggregated states. Thus, this modified phospholipid system and its simple grafting technique was very effective with regard to suppressing in vitro platelet adhesion on the silicon catheter surface.