Suitability of the German test method for cleaning efficacy in washer-disinfectors for flexible endoscopes according
B Zühlsdorf1, G Kampf, H Floss
1Technische Hygiene, Charité-Universitätsmedizin Berlin, Campus Benjamin Franklin, Hindenburgdamm 27, 12203 Berlin, Germany.
Abstract:
In Europe, the evaluation of processing flexible endoscopes in washer-disinfectors (WDs) is performed in compliance with prEN ISO 15883-1 which includes determination of the efficacy of the cleaning process. Recent data suggest that cleaning processes show large differences when the prEN ISO 15883-1 German test model is applied. Hence, we analysed a total of 72 experiments in order to evaluate the test method. Transparent test tubes as test pieces (length 2 m, lumen 2 mm) were contaminated with a mixture of blood and Enterococcus faecium. Three set-ups were used: WD 425 with soft water, WD 425 with hard water and WD 440 with demineralized water. WDs were set to perform the cleaning stage of the programme alone. Seven cleaning agents were used according to the manufacturers' instructions (21 cleaning processes); in addition, three cleaning processes were carried out without a cleaning agent (i.e. with water alone). Each cleaning process was assessed by means of three experiments. Suspensions of test organism had 9.2x10(10) colony-forming units (cfu)/mL E. faecium (mean of 24 processes). Controls (recovery) contained 1.0x10(6) cfu/mL E. faecium (mean of 71 experiments). Mean log(10) reduction factors (RFs) for each process, i.e. the difference in microbial loads on the control and the processed tubes, were calculated. Cleaning processes led to RFs of 0-4.1, but no process led to residual bacterial loads below the limit of detection (1.8l gcfu/mL). Standard deviations for a cleaning process were small (< or =0.6 in 79% of the processes) indicating adequate reproducibility. The test model led to reproducible results and revealed large differences between the individual processes. If a cleaning process is intended to result in a bioburden reduction (i.e. RF> or =4), the control must carry a minimum bioburden of 6.5x10(5) cfu/mL. This was achieved in 58% of the processes. However, controls with a bioburden <6.5x10(5) cfu/mL never yielded a residual bacterial load below the limit of detection. We found that the prEN ISO 15883-1 German test method is suitable to determine the cleaning efficacy in WDs and leads to reproducible and valid results.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...