Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
Separation and detection of phosphorylated and nonphosphorylated peptides in liquid chromatography-mass spectrometry
Andreas Tholey1, Hansjörg Toll, Christian G Huber
1Technische Biochemie and Instrumentelle Analytik und Bioanalytik, Saarland University, 66123 Saarbrücken, Germany. a.tholey@mx.uni-saarland.de
Abstract:
Efficient chromatographic separation is a prerequisite for the sensitive analysis of complex peptide mixtures using liquid chromatography-mass spectrometry. This is especially true for the analysis of mixtures of unmodified and posttranslationally modified peptides, for example, phosphorylated peptides in the presence of their unmodified analogues. Applying monolithic capillary columns based on poly(styrene/divinylbenzene), the influence of acidic eluents based on trifluoroacetic and heptafluorobutyric acid as well as an alkaline eluent based on triethylamine-acetic acid (pH 9.2) on the separation of synthetic phosphopeptides was evaluated. Heptafluorobutyric acid offered the longest retention times and highest selectivities and, hence, the most effective separation. Application of the alkaline eluent in conjunction with detection in negative ion mode electrospray ionization mass spectrometry, on the other hand, allowed the detection of phosphorylated peptides with significantly lower limits of detection, as compared to acidic eluents in combination with detection in positive ion mode. Pairs of phosphorylated and nonphosphorylated synthetic peptides, ranging from 7- to 16-mers, as well as phosphorylated peptides form a tryptic protein digest could be separated both at acidic and alkaline pH. Utilizing a 60 x 0.20-mm-i.d. capillary column, the limit of detection in negative ion detection mode for a 4-fold phosphorylated peptide in a beta-casein digest was 10 fmol. Together with the capability for fast separation of protein digests, monolithic columns, thus, facilitate the effective and sensitive analysis of this important posttranslational modification.
Related Concept Videos
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Types of Detectors
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Principles Of Column Chromatography
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...