Related Experiment Video
Updated: Aug 16, 2026

Measurement and Analysis of Atomic Hydrogen and Diatomic Molecular AlO, C2, CN, and TiO Spectra Following Laser-induced Optical Breakdown
Published on: February 14, 2014
Joint analyses by laser-induced breakdown spectroscopy (LIBS) and Raman spectroscopy at stand-off distances
Roger C Wiens1, Shiv K Sharma, Justin Thompson
1Space Science and Applications, Mail Stop D-466, Los Alamos National Laboratory, Los Alamos, NM 87544, USA. rwiens@lanl.gov
Abstract:
Raman spectroscopy and laser-induced breakdown spectroscopy (LIBS) of solid samples have both been shown to be feasible with sample-to-instrument distances of many meters. The two techniques are very useful together, as the combination of elemental compositions from LIBS and molecular vibrational information from Raman spectroscopy strongly complement each other. Remote LIBS and Raman spectroscopy spectra were taken together on a number of mineral samples including sulfates, carbonates and silicates at a distance of 8.3 m. The complementary nature of these spectra is highlighted and discussed. A factor of approximately 20 difference in intensity was observed between the brightest Raman line of calcite, at optimal laser power, and the brighter Ca I LIBS emission line measured with 55 mJ/pulse laser power. LIBS and Raman spectroscopy have several obstacles to devising a single instrument capable of both techniques. These include the differing spectral ranges and required detection sensitivity. The current state of technology in these areas is discussed.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to the...
UV–Vis Spectroscopy: Beer–Lambert Law
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...