Related Experiment Video
Updated: Aug 16, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Comparison of the intracuff pressures of three different tracheostomy tubes
1Department of Anesthesiology, Faculty of Medicine, The University of Tokyo, Tokyo, Japan.
Abstract:
The purpose of this study was to compare the cuff pressures of three tracheostomy tubes, MERA sofit CLEAR, Blue Line Tracheostomy Tube, and Tracheosoft. Each tracheostomy tube with an internal diameter of 7.0 mm was put into a plastic column. The cuff was then inflated with air to seal the column, and the column was filled with water. The air in the cuff was withdrawn gradually and the cuff pressure at the point of water leakage was measured. Six columns of different size were used. In columns with an internal diameter of 18-21 mm, the water leakage pressure was lower in the following order: MERA sofit CLEAR < Tracheosoft < or = Blue Line Tracheostomy Tube. A mongrel dog was tracheotomized, and each tracheostomy tube with an internal diameter of 7.5 mm was intubated. The cuff air was increased by 1 ml from 4 ml to 10 ml, and the intracuff pressure was then measured. The intracaff pressure of the Blue Line Tracheostomy Tube was the highest at the same cuff volume, and that of the Tracheosoft was next. Based on these results, the MERA sofit CLEAR was found to maintain most safely the lowest intracuff pressure to seal the trachea among the three tracheostomy tubes tested.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...