Related Experiment Video
Updated: Aug 16, 2026

Corneal Donor Tissue Preparation for Endothelial Keratoplasty
Published on: June 12, 2012
Amadeus microkeratome: experience with the first 2000 cases and lessons learned
Colin Ck Chan1, Con N Moshegov
1Department of Ophthalmology, Concord Hospital, Sydney, New South Wales, Australia.
Background:
To assess the safety and efficacy of the Amadeus microkeratome and to report on lessons learnt in optimizing results from its use.
Methods:
A retrospective review of the flap-related complications of 2000 laser in situ keratomileusis (LASIK) procedures performed by one surgeon using the Amadeus microkeratome (AMO, Irvine, CA, USA) was conducted. One hundred consecutive cases had preoperative and intraoperative pachymetry to evaluate the relationship between flap thickness, corneal thickness and reuse of the blade.
Results:
There were 35 cases of minor epithelial defects (1.8%) and four eyes with major epithelial defects (0.2%). There were no cases of suction loss, buttonholes, or incomplete or damaged flaps. There were four free caps (0.2%). Twenty-five cases had minor wrinkles and two cases had significant folds requiring relifts. Thinner corneas resulted in thinner flaps (P < 0.0001). Reuse of the blade resulted in thinner flaps (P = 0.0001).
Conclusion:
The Amadeus is a safe and reliable microkeratome. Although short, there is a learning curve with use of the instrument. Corneal thickness and reuse of the blade affect flap thickness.