A pilot study of continuous transtracheal mixed venous oxygen saturation monitoring
Wei Wei1, Zhaoqiong Zhu, Lunxu Liu
1Departments of Anesthesiology and Cardiothoracic Surgery, West China Hospital, Sichuan University; Department of Physics, Sichuan University, Chengdu, Sichuan, P. R. China; and the Department of Anesthesia, Plastic Surgery Hospital, Chinese Academy of Medical Sciences, Beijing, P. R. China.
Unlabelled:
In this study, we investigated the feasibility and the accuracy of transtracheal mixed venous oxygen saturation (Svo(2)) monitoring. Ten patients undergoing thoracic surgery were included in this study. A single-use pediatric pulse oximetry sensor was attached to the double-lumen tube between the tracheal and bronchial cuff. After anesthesia was induced, the double-lumen tube was inserted into the trachea and adjusted to the proper position. During surgery, the pulmonary arterial blood was sampled every 3 min for 15 min to measure the Svo(2). The measurements made by the transtracheal pulmonary pulse oximeter (Sto(2)) were recorded at the same time that blood was sampled from the pulmonary artery for Svo(2) measurements. The levels of measurement agreement between the Sto(2) and the Svo(2) were analyzed using the Bland and Altman method. The mean +/- sd (range) oxygen saturation values during the data collecting period were 82.0% +/- 4.9% (72%-91%) for the Sto(2) and 82.2% +/- 5.5% (71%-91%) for the Svo(2), respectively. The linear correlation coefficient of the regression analysis between the Sto(2) and the Svo(2) was 0.934 (P < 0.05). A 95% confidence interval for absolute difference between the Sto(2) and the Svo(2) was 1.58%-2.09%. The mean +/- 2 sd difference between the Sto(2) and the Svo(2) was 0.12% +/- 3.97% on the Bland and Altman graph. We conclude that it is feasible to monitor the pulmonary artery oxygen saturation continuously by a transtracheal pulse oximetry technique and that it can be done so accurately.
Implications:
Mixed venous oxygen saturation (Svo2) is a measure of the balance between oxygen supply and consumption throughout the whole body. Svo2 can be measured invasively by inserting a pulmonary artery catheter with the associated disadvantages of cost and potential for patient injury. In this study, we investigated the feasibility of noninvasive Svo2 measurement using a transtracheal pulse oximetry technique.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...