Related Experiment Video
Updated: Aug 16, 2026

Semi-Targeted Ultra-High-Performance Chromatography Coupled to Mass Spectrometry Analysis of Phenolic Metabolites in Plasma of Elderly Adults
Published on: April 22, 2022
Autoanalyzer for continuous fractionation and quantitation of the polyphenols content in wines
Rafael Lucena1, Soledad Cárdenas, Mercedes Gallego
1Department of Analytical Chemistry, Marie Curie Building (Annex), Campus de Rabanales, University of Córdoba, E-14071 Córdoba, Spain. qa1meobj@uco.es
Abstract:
A simple continuous flow autoanalyzer for the on-line fractionation of the polyphenols content in wines is proposed. The target compounds are isolated from the matrix by solid-phase extraction on an RP-C18 sorbent column, using selective solvents for the sequential elution of each polyphenol family. Moreover, evaporative light scattering detection (ELSD) is used for the first time for the on-line monitorization of the three polyphenol fractions present in the wine samples. Thus, a single sample injection is required to determine the global concentration of the three selected polyphenol fractions and the whole analysis is completed within a few minutes. Three calibration graphs were constructed for quantitative analysis of the global compounds concentration in every fraction and covered the range 5-300 mg l(-1) (expressed as gallic acid). Average repeatability, expressed as relative standard deviation, was 4%. The proposed autonalyzer was applied to the analysis of a variety of commercial wine samples. The results obtained were compared with those provided by the Folin-Ciocalteau method, being similar in all instances.