Related Experiment Videos
Radiation exposure to the hands from mini C-arm fluoroscopy
1Department of Orthopedics, Kaiser Permanente, Denver, CO 80205, USA. gordobrig@aol.com
The Journal of Hand Surgery
|July 26, 2005
Summary
Hand surgeons experience significant radiation exposure to their hands during mini C-arm fluoroscopy, averaging 20 mrem per case. This highlights the need for protective measures to minimize hand radiation exposure during these procedures.
Area of Science:
- Orthopedic Surgery
- Medical Imaging
- Radiation Safety
Background:
- Intraoperative fluoroscopy using mini C-arms is common in hand surgery.
- Radiation exposure to surgical staff, particularly hands, is a concern.
- Understanding exposure levels is crucial for implementing safety protocols.
Purpose of the Study:
- To quantify hand radiation exposure for surgeons using intraoperative mini C-arm fluoroscopy.
- To compare measured exposure levels with predicted values and safety limits.
- To assess the risk of radiation exposure during finger, hand, and wrist surgeries.
Main Methods:
- Ring dosimeters were used to measure hand radiation exposure in five hand surgeons during 81 surgeries.
- Fluoroscopy time per case was recorded.
- A phantom study was conducted to measure radiation scatter at close range.
Main Results:
- Surgeons' hands received an average of 20 +/- 12.3 mrem of radiation per case.
- Individual exposure ranged from 5 to 80 mrem per case.
- Average fluoroscopy time was 51 +/- 36.9 seconds, with a range of 6 to 170 seconds, suggesting hands may enter the direct beam.
Conclusions:
- Hand surgeons face substantial radiation exposure due to proximity to the mini C-arm beam.
- The findings indicate a need for enhanced radiation safety precautions for hand surgeons.
- Minimizing direct beam exposure is recommended to reduce hand radiation levels.