The "Lasso-o" tape: stretchability and observer variability in head circumference measurement
J L Bartram1, A S Rigby, P S Baxter
1Sheffield Children's Hospital, UK. jacklbartram@hotmail.com
Archives of Disease in Childhood
|July 26, 2005
Summary
The Lasso-o tape, used for head circumference measurements in children, stretches significantly over time. While generally reproducible, some measurements show over 1 cm variability between observers.
Area of Science:
- Pediatric health
- Clinical measurement tools
- Biomedical engineering
Background:
- Accurate head circumference measurement is vital for pediatric health assessment.
- The Lasso-o tape is a commonly used instrument for this measurement.
- Understanding tape properties and measurement variability is crucial for reliable clinical data.
Purpose of the Study:
- To evaluate the stretchability of the Lasso-o tape.
- To assess the intra- and inter-observer reproducibility of measurements obtained with the Lasso-o tape.
- To identify potential sources of measurement error in pediatric head circumference assessment.
Main Methods:
- Investigated the physical properties of the Lasso-o tape, specifically its stretchability.
- Conducted repeated measurements of head circumference using the Lasso-o tape by single observers (intra-observer) and multiple observers (inter-observer).
- Calculated intra-class coefficients (ICC) to quantify measurement reproducibility.
Main Results:
- Significant stretching was observed in older Lasso-o tapes.
- High intra-observer reproducibility was found, with an ICC of 0.999.
- Good inter-observer reproducibility was observed (ICC = 0.979), but 9% of measurements exhibited variability exceeding 1 cm.
Conclusions:
- The Lasso-o tape's stretchability, particularly in older tapes, can impact measurement accuracy.
- While generally reliable, inter-observer variability exceeding 1 cm in a small percentage of cases warrants attention.
- Clinical guidelines should consider tape condition and observer training to minimize measurement errors in pediatric head circumference assessment.