A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Comparison of different body size parameters for individual dose adaptation in body CT of adults
1Department of Diagnostic Radiology, University Hospital, Rosenwinkel 5, 37081 Goettingen, Germany. Menke_J@T-Online.de
Purpose:
To investigate prospectively which of several body size parameters are suitable for individual dose adaptation in body computed tomography (CT) of adults.
Materials And Methods:
Three body regions (thorax, abdomen, pelvis) were scanned exclusively for clinical reasons, with institutional ethical approval and informed consent. For each of the three regions, 50 men and 50 women (aged 18-87 years) were studied (300 scans total). Individual x-ray properties for each scan were summarized with a water-equivalent diameter (Dw). Different body size parameters, based on weight, height, and shape, were correlated with Dw by using regression analysis. This resulted in Dw estimation errors of different magnitudes, indicated with 95% prediction intervals. The errors from weight were compared with those from each of the other body parameters by using comparison of variance in paired samples (P < .05). In addition, a topogram-based estimate for Dw was studied, which simulated an automated body size measurement.
Results:
For the thorax, abdomen, and pelvis, mean Dw was 28.0, 29.1, and 29.3 cm, and estimation of Dw from weight enabled 95% prediction intervals of +/-2.5, +/-2.4, and +/-2.6 cm, respectively. Combinations of height and weight were only slightly more or even less exact than were measurements from only weight. Diameter-related parameters such as body circumference were similar to or better than weight. However, the topogram-based estimate was significantly more exact.
Conclusion:
Body weight and circumference enable suitable estimates for individual dose adaptation in body CT of adults if automated dose adaptation is not available.
Related Concept Videos
Dosage Regimen: Individualization
Drug Dosing: Infants and Children
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation