Related Experiment Video
Updated: Aug 16, 2026

Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015
A thermochromic dispersive electrode can measure the underlying skin temperature and prevent burns during
Aravinda Thiagalingam1, Jim Pouliopoulos, Michael Anthony Tony Barry
1Department of Cardiology, Westmead Hospital, Westmead, Australia. aravinda@westgate.wh.usyd.edu.au
Unlabelled:
Evaluation of a thermochromic dispersive electrode.
Introduction:
Burns at the dispersive electrode are serious complications of diathermy and radiofrequency (RF) ablation procedures. We aimed to create a new methodology to reduce the incidence of dispersive electrode related skin burns. We hypothesized that a dispersive electrode incorporating a thermochromic liquid crystal (TLC) layer could accurately measure underlying skin temperatures and help prevent burns.
Methods And Results:
The TLC electrode was compared with a standard dispersive electrode in 12 male sheep. RF current was delivered with the dispersive electrode fully applied or partially detached to simulate different clinical scenarios. The temperature of the TLC layer, calculated from the hue (color) every 15 seconds, was compared with fluoroptic skin temperature probes. TLC electrodes with a temperature range of 45-58 degrees C were used in six sheep to assess the correlation of TLC temperature distribution with skin temperature and burns. TLC electrodes with a temperature range of 40-50 degrees C were used in another 6 sheep to simulate clinical conditions in which the ablation was stopped if the TLC temperature was >42 degrees C. The TLC measured temperatures correlated well with fluoroptic probes at the skin surface (r=0.94+/-0.05, mean of the absolute difference in temperature difference 0.9+/-0.58 degrees C). Ablations with partial application of standard dispersive electrodes consistently caused skin burns. There were no burns under the TLC electrode when ablations were ceased for temperatures>42 degrees C.
Conclusions:
TLC-equipped dispersive electrodes were able to accurately measure skin temperature under the electrode. This technology is likely to prevent dispersive electrode related burns.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
05:43Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Thermosensation
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Decreased Body Temperature
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...