How well do clinic-based blood pressure measurements agree with the mercury standard?
Jennifer W Kim1, Hayden B Bosworth, Corrine I Voils
1Duke University School of Medicine, Durham, NC, USA.
Background:
Obtaining accurate blood pressure (BP) readings is a challenge faced by health professionals. Clinical trials implement strict protocols, whereas clinical practices and studies that assess quality of care utilize a less rigorous protocol for BP measurement.
Objective:
To examine agreement between real-time clinic-based assessment of BP and the standard mercury assessment of BP.
Design:
Prospective reliability study.
Patients:
One hundred patients with an International Classification of Diseases-9th edition code for hypertension were enrolled.
Measures:
Two BP measurements were obtained with the Hawksley random-zero mercury sphygmomanometer and averaged. The clinic-based BP was extracted from the computerized medical records.
Results:
Agreement between the mercury and clinic-based systolic blood pressure (SBP) was good, intraclass correlation coefficient (ICC)=0.91 (95% confidence interval (CI): 0.83 to 0.94); the agreement for the mercury and clinic-based diastolic blood pressure (DBP) was satisfactory, ICC=0.77 (95% CI: 0.62 to 0.86). Overall, clinic-based readings overestimated the mercury readings, with a mean overestimation of 8.3 mmHg for SBP and 7.1 mmHg for DBP. Based on the clinic-based measure, 21% of patients were misdiagnosed with uncontrolled hypertension.
Conclusions:
Health professionals should be aware of this potential difference when utilizing clinic-based BP values for making treatment decisions and/or assessing quality of care.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.