The NerveSeeker: a system for automated nerve localization.

S A Raymond1, S B Abrams, D B Raemer

  • 1Department of Anesthesia, Brigham and Women's Hospital, Boston, MA 02115.

Regional Anesthesia
|May 1, 1992
PubMed
Summary

The NerveSeeker, a novel nerve stimulator, automatically adjusts current to precisely locate peripheral nerves. This instrument successfully tracked needle movements, indicating readiness for clinical trials.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Using screen-based simulation of inhaled anaesthetic delivery to improve patient care.

British journal of anaesthesia·2015
Same author

Bupivacaine inhibits endothelin-1-evoked increases in intracellular calcium in model sensory neurons.

Acta anaesthesiologica Scandinavica·2015
Same author

State-Dependent Inhibition of Sodium Channels by Local Anesthetics: A 40-Year Evolution.

Biochemistry (Moscow) Supplement. Series A, Membrane and cell biology·2013
Same author

Successful treatment of hydramnios with ammonium chloride; a preliminary report.

American journal of obstetrics and gynecology·2010
Same author

Contralateral paw sensitization following injection of endothelin-1: effects of local anesthetics differentiate peripheral and central processes.

Neuroscience·2009
Same author

Novel ideas of local anaesthetic actions on various ion channels to ameliorate postoperative pain.

British journal of anaesthesia·2008

Area of Science:

  • Biomedical Engineering
  • Neuroscience
  • Medical Devices

Background:

  • Peripheral nerve blocks are crucial for anesthesia and pain management.
  • Accurate nerve localization is essential to improve block efficacy and minimize complications.
  • Current nerve stimulators require manual adjustment of stimulating current.

Purpose of the Study:

  • To evaluate the performance of the NerveSeeker, a new nerve stimulator with automatic current adjustment.
  • To assess the NerveSeeker's ability to maintain a constant neural response during needle insertion.
  • To determine optimal parameters for the NerveSeeker's feedback control system.

Main Methods:

  • Utilized an excised frog nerve in a transparent chamber for direct observation.
  • Employed a feedback control system to automatically adjust stimulating current amplitude.

Related Experiment Videos

  • Correlated control voltage with stimulus magnitude to indicate needle proximity to the nerve.
  • Analyzed device performance in tracking needle movements at various velocities.
  • Main Results:

    • A combined strategy of reducing stimulus on approach and incrementing on withdrawal proved effective.
    • The NerveSeeker maintained a constant neural response, guiding needle placement.
    • Nerve localization was successful, with the device tracking needle movements up to 2 mm/second.
    • Optimal control parameters were identified, balancing engineering and physiological factors.

    Conclusions:

    • Experimental results demonstrate the NerveSeeker's efficacy in peripheral nerve localization.
    • The device's automatic current adjustment offers improved guidance during nerve blocks.
    • Clinical testing of the NerveSeeker is warranted based on successful experimental validation.