Visual simulation of retinal images through a decentered monofocal and a refractive multifocal intraocular lens
Kazuno Negishi1, Kazuhiko Ohnuma, Takashi Ikeda
1Department of Ophthalmology, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo 160-8582, Japan. fwic7788@infoweb.ne.jp
Purpose:
To evaluate the effect of decentration of a monofocal intraocular lens (IOL) and a refractive multifocal IOL on retinal image quality using a new visual simulation system.
Methods:
Using a new visual simulation system, we performed visual simulation of a monofocal and a multifocal IOL at 5, 4, 3, 2, 1, and 0.4 m with several decentered IOL positions from 0 to 1.0 mm through a 3- or 4-mm aperture using Landolt visual acuity (VA) charts. The VA was estimated under each condition from the simulated retinal image.
Results:
With a monofocal IOL, the image was affected minimally by decentration at 4 and 5 m; at 2 and 3 m, the image contrast decreased slightly with increased decentration. With the multifocal IOL, some loss of image contrast developed at all distances compared with the monofocal IOL; however, the images of the Landolt's rings were still recognizable under all conditions.
Conclusions:
Our results suggest that up to 1.0 mm of decentration of a monofocal and multifocal IOL would not greatly affect the retinal image quality.