Chemiluminescence biosensor system for lactic acid using natural animal tissue as recognition element
Fangqiong Wu1, Yuming Huang, Chengzhi Huang
1College of Chemistry and Chemical Engineering, Chongqing Key Laboratory of Modern Analytical Sciences, Southwest Normal University, Beibei, Chongqing 400715, PR China.
Biosensors & Bioelectronics
|August 4, 2005
Summary
This study introduces a novel chemiluminescence biosensor using porcine kidney tissue for detecting lactic acid. The developed method offers a sensitive, rapid, and stable approach for analyzing lactic acid in biological samples.
Area of Science:
- Biotechnology
- Analytical Chemistry
- Biochemistry
Background:
- Lactic acid is a crucial biomarker in various biological and food samples.
- Accurate and rapid detection of lactic acid is essential for clinical diagnostics and quality control.
- Existing methods for lactic acid detection may have limitations in sensitivity, speed, or cost.
Purpose of the Study:
- To develop a novel chemiluminescence biosensor for lactic acid detection.
- To utilize natural animal tissue (porcine kidney) as a recognition element.
- To evaluate the performance and applicability of the developed biosensor.
Main Methods:
- A biosensor was constructed using porcine kidney tissue containing alpha-hydroxy acid oxidase.
- Lactic acid was oxidized to produce hydrogen peroxide, which reacted with luminol and potassium ferricyanide to generate a chemiluminescence (CL) signal.
- The CL signal intensity was measured and correlated with lactic acid concentration.
Main Results:
- The CL emission intensity showed a linear response to lactic acid concentration in the range of 1-1000 micromol/L.
- A low detection limit of 0.2 micromol/L (3sigma) was achieved for lactic acid.
- The biosensor demonstrated continuous usability for 6 hours with over 240 measurements and high reproducibility (RSD < 5%).
- Analysis time per sample was as short as 1.5 minutes with a relative standard deviation of 1.12% for 100 micromol/L lactic acid.
Conclusions:
- The developed porcine kidney-based biosensor is a sensitive, rapid, and stable platform for lactic acid detection.
- The method is suitable for the quantitative analysis of lactic acid in complex matrices like plasma and milk.
- This approach offers a promising alternative for routine lactic acid monitoring.