Fluorescent light effects on FWT-60 radiochromic film
Martin J Butson1, Tsang Cheung, Peter K N Yu
1Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong, Hong Kong. butsonm@iahs.nsw.gov.au
Abstract:
FWT-60 radiochromic film has been tested for colouration effects from fluorescent light sources and shown to produce a marked colouration when exposed to office fluorescent light sources showing an approximate 1 OD unit per 0.5 J m(-2) exposure to a broad ultraviolet (UV) UVA + UVB spectrum at the peak absorption wavelength. This produces a measurable and quantifiable response to UV exposure. By choosing an appropriate wavelength of readout or band pass, the level of sensitivity can be changed to match the application or exposure level measurement required. These levels of UV response are significantly higher in sensitivity than other radiochromic films such as Gafchromic MD-55 by an order of magnitude. This feature may be of use for measurement of integrated UV exposure from fluorescent lights when required and produces a quantifiable history of total exposure.