Related Experiment Video
Updated: Aug 16, 2026

Enhanced Cochlear Coverage and Hearing Preservation in High-Frequency Hearing Loss via Electric Acoustic Stimulation with Longer Electrode
Published on: October 11, 2024
Acoustic simulations of combined electric and acoustic hearing (EAS)
Michael F Dorman1, Anthony J Spahr, Philipos C Loizou
1Department of Speech and Hearing Science, Arizona State University, Tempe, Arizona 85287-0102, USA.
Objective:
Our aim was to explore the consequences for speech understanding of leaving a gap in frequency between a region of acoustic hearing and a region stimulated electrically. Our studies were conducted with normal-hearing listeners, using an acoustic simulation of combined electric and acoustic (EAS) stimulation.
Design:
Simulations of EAS were created by low-pass filtering speech at 0.5 kHz (90 dB octave roll-off) and adding amplitude-modulated sine waves at higher frequencies. The gap in frequency between acoustic and simulated electric hearing was varied over the range 0.5 kHz to 3.2 kHz. Stimuli included sentences in quiet, sentences in noise, and consonants and vowels. Three experiments were conducted with sample sizes of 12 listeners.
Results:
Scores were highest in conditions that minimized the frequency gap between acoustic and electric stimulation. In quiet, vowels and consonant place of articulation showed the most sensitivity to the frequency gap. In noise, scores in the simulated EAS condition were higher than the sum of the scores from the acoustic-only and simulated electric-only conditions.
Conclusions:
Our results suggest that both deep and shallow insertions of electrodes could improve the speech understanding abilities of patients with residual hearing to 500 Hz. However, performance levels will be maximized if the gap between acoustic and electric stimulation is minimized.
More Related Videos
09:18A Comparative Study of Drug Delivery Methods Targeted to the Mouse Inner Ear: Bullostomy Versus Transtympanic Injection
Published on: March 8, 2017
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
The Cochlea
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case, then the...
Anatomy of the Ear