Related Experiment Video
Updated: Aug 16, 2026

Probing The Structure And Dynamics Of Nucleosomes Using Atomic Force Microscopy Imaging
Published on: January 31, 2019
Formation of aminosilane-functionalized mica for atomic force microscopy imaging of DNA
Neal Crampton1, William A Bonass, Jennifer Kirkham
1School of Physics and Astronomy, University of Leeds, Leeds LS2 9JT, UK.
Abstract:
Factors affecting the functionalization of mica with aminosilanes, in particular, aminopropyltriethoxysilane (APTES-mica), formed from the vapor phase have been systematically studied. The relative humidity (RH) during vapor deposition has been varied, and postdeposition treatment through baking has been used, as well as the comparison of mono and trifunctionality, to investigate how optimal surfaces for AFM imaging of DNA are formed. It is found that the stability of the APTES layers is a consequence of lateral polymerization and not covalent attachment to the mica substrate. At low RH (<25%), DNA adopts an open, well-resolved conformation, whereas at >25% RH, DNA surface-induced condensation occurs. Contact mode AFM scratching experiments show that two main structures of the silane layer exist at different humidity: a monolayer exists at RH < 25%, and a bilayer structure exists at RH > 25%. Finally, structural changes that these two layer types undergo after baking at 150 degrees C were investigated by AFM and X-ray photoelectron spectroscopy (XPS), and these now prevented DNA from binding to the APTES-mica, except in the presence of Mg(II) ions.