Increased sensitivity of bacterial detection in cerebrospinal fluid by fluorescent staining on low-fluorescence
Jacob D Durtschi1, Maria Erali1, L Kathryn Bromley1
1ARUP Institute for Clinical and Experimental Pathology1 and Department of Pathology2, University of Utah, Salt Lake City, UT 84108, USA.
Abstract:
A membrane-filter-based, fluorescent Gram stain method for bacterial detection in cerebrospinal fluid samples was developed and evaluated as a rapid, sensitive alternative to standard Gram stain protocols. A recently developed, modified version of the aluminium oxide membrane Anopore with low-fluorescence optical properties showed superior performance in this application. Other aspects of the fluorescent Gram stain system that were evaluated include membrane filter selection, strategies to reduce fluorescence fading and the effect of patient blood cells on bacterial detection in the fluorescently stained cerebrospinal fluid samples. The combination of the membrane filter's bacteria-concentrating ability and absolute retention along with high-contrast, fluorescent Gram discriminating dyes enabled rapid bacterial detection and Gram discrimination, with a 1-1.5 order of magnitude increase in the bacterial concentration limit of detection.
More Related Videos
09:24Optical Screening of Novel Bacteria-specific Probes on Ex Vivo Human Lung Tissue by Confocal Laser Endomicroscopy
Published on: November 29, 2017
11:17Monitoring Changes in Membrane Polarity, Membrane Integrity, and Intracellular Ion Concentrations in Streptococcus pneumoniae Using Fluorescent Dyes
Published on: February 17, 2014