Nitrous oxide diffusion into tracheal tube cuffs--efficacy of a new prototype cuff pressure release valve
A Dullenkopf1, A C Gerber, M Weiss
1Department of Anaesthesia, University Children's Hospital, Zurich, Switzerland. alex.dullenkopf@kispi.unizh.ch
Background:
The aim of this study was to evaluate the performance of a new cuff pressure release valve (CPRV), in which the release pressure can be adjusted from 10 to 25 cmH2O, particularly intended to control pressure in paediatric cuffed tracheal tubes and to avoid cuff hyperinflation caused by N2O diffusion.
Methods:
In vitro: the PRV was set to 10, 15, 20 or 25 cmH2O release pressure and connected to a cuffed tube placed into a box flushed with 66% N2O in O2. The cuff pressure was monitored with and without CPRV for 60 min. Experiments were performed four times using two different CPRVs. In vivo: with Institutional Review Board approval, CPRV was studied in 50 children undergoing general anaesthesia with tracheal intubation and standardized anaesthesia technique (including 66% N2O in O2) and ventilator settings. Patients were randomized into two groups (with and without CPRV). The cuff pressure baseline was 20 cmH2O and CPRV was set to 25 cmH2O. If the cuff pressure exceeded 25 cmH2O, it was manually released to 20 cmH2O. The numbers of deflations in both groups were noted and compared by Mann-Whitney U-test (P<0.05).
Results:
In vitro: the cuff pressure exceeded 50 cmH2O after 60 min without CPRV, but did not exceed the settings with CPRV. In vivo: there was no need to manually deflate the cuff in the CPRV group but, in every patient in the control group, three (two to seven) deflating manoeuvres were required within the first hour of anaesthesia (P<0.0001).
Conclusion:
The CPRV allows reliable cuff pressure release at various pressure levels and reliably prevents cuff pressure increases caused by N2O.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)