Related Experiment Video
Updated: Aug 16, 2026

A Modified QuEChERS-HPLC Method for Detection of Polycyclic Aromatic Hydrocarbons in Zebrafish Embryos Exposed to Fine Particulate Matter
Published on: June 13, 2025
Ionic liquid-containing semipermeable membrane devices for monitoring the polycyclic aromatic hydrocarbons in water
Wenyan Zhao1, Meng Han, Shugui Dai
1College of Chemistry and Chemical Engineering, Inner Mongolia University, Hohhot 010021, China.
Abstract:
Ionic liquid-containing semipermeable membrane devices (IL-SPMDs) were developed to monitor the polycyclic aromatic hydrocarbons in water. Uptake kinetics of naphthalene, 1-methylnaphthalene, phenanthrene, pyrene, chrysene by layflat low-density polyethylene tubing (15 cm x 2 cm) filled with 0.5 ml 1-butyl-3-methylimidazolium hexafluorophosphate ionic liquid were studied in a laboratory continuous-flow system for the duration of 16 days. The device concentration factors were in the range of 830-7800 for the studied PAHs. The kinetic parameters of IL/water partition coefficients, analytes uptake rate constants and sampling rates were estimated using mathematical models. These parameters were used in the field experiment to estimate the concentrations of the PAHs in Lanzhou section of Yellow River in China, with the result in comparison with that obtained by triolein-SPMD.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Potentiometry: Membrane Electrodes
High-Performance Liquid Chromatography: Types of Detectors
Dialysis