Failure of gated single photon emission computer tomography scan to detect imminent acute plaque rupture causing
1Nucler Cardiology, University of California, Irvine Medical Center, 101 The City Dr, Bldg 53, Rm 100, Orange, CA 92868-4080, USA. mmovahed@uci.edu
Abstract:
The negative predictive value of a gated single photon emission computer tomography (SPECT) scan is very high, with an event rate of < 1% in the first year. However, the presence of nonobstructive coronary artery plaque should yield normal SPECT scan findings. On the other hand, most plaque ruptures, which are a major cause of acute myocardial infarction, occur in nonobstructive coronary artery plaque. Therefore, the findings of a gated SPECT scan should be normal if a ruptured plaque has not created significant obstruction despite the imminent threat of coronary artery occlusion. We present the first case report of a documented gated SPECT scan in a patient who had experienced an acute anterior Q-wave myocardial infarction showing no significant ischemia in the anterior wall in the last minute of data acquisition.