The effect of source to image-receptor distance on effective dose for some common X-ray projections
1Unitec New Zealand, Private Bag 92025, Auckland, New Zealand.
Abstract:
A number of studies on the effects of source to image distance (SID) on dose to the patient have been published. These generally assume that the X-ray beam is collimated to a rectangular region of fixed size at the entrance surface of the patient at each SID. This is poor radiographic practice. In this work, effective doses have been determined using the commercially available Monte Carlo simulation package PCXMC. Three collimation cases were considered, namely to regions of clinical interest at the entrance and at the centre of the patient, and a fixed beam size at the image receptor. For cases of collimation to a central region of interest or to the image receptor, the saving in effective dose at increased SIDs is modest.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...