Determination of CT-to-density conversion relationship for image-based treatment planning systems
Cheng B Saw1, Alphonse Loper, Krishna Komanduri
1Department of Radiation Oncology, University of Pittsburgh Medical Center Cancer Centers, Pittsburgh, PA 15232, USA. cbsaw@yahoo.com
Abstract:
The implementation of tissue inhomogeneity correction in image-based treatment planning will improve the accuracy of radiation dose calculations for patients undergoing external-beam radiotherapy. Before the tissue inhomogeneity correction can be applied, the relationship between the computed tomography (CT) value and density must be established. This tissue characterization relationship allows the conversion of CT value in each voxel of the CT images into density for use in the dose calculations. This paper describes the proper procedure of establishing the CT value to density conversion relationship. A tissue characterization phantom with 17 inserts made of different materials was scanned using a GE Lightspeed Plus CT scanner (120 kVp). These images were then downloaded into the Eclipse and Pinnacle treatment planning systems. At the treatment planning workstation, the axial images were retrieved to determine the CT value of the inserts. A region of interest was drawn on the central portion of the insert and the mean CT value and its standard deviation were determined. The mean CT value was plotted against the density of the tissue inserts and fitted with bilinear equations. A new set of CT values vs. densities was generated from the bilinear equations and then entered into the treatment planning systems. The need to obtain CT values through the treatment planning system is very clear. The 2 treatment planning systems use different CT value ranges, one from -1024 to 3071 and the other from 0 to 4096. If the range is correct, it would result in inappropriate use of the conversion curve. In addition to the difference in the range of CT values, one treatment planning system uses physical density, while the other uses relative electron density.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography