Validity and reproducibility of arterial pulse wave velocity measurement using new device with oscillometric
Madireddy Umamaheshwar Rao Naidu1, Budda Muralidhar Reddy, Sridhar Yashmaina
1Department of Clinical Pharmacology & Therapeutics, Nizam's Institute of Medical Sciences, Punjagutta, Hyderabad, 500082, India. murnaidu@yahoo.com
Background:
Availability of a range of techniques and devices allow measurement of many variables related to the stiffness of large or medium sized arteries. There is good evidence that, pulse wave velocity is a relatively simple measurement and is a good indicator of changes in arterial properties. The pulse wave velocity calculated from pulse wave recording by other methods like doppler or tonometry is tedious, time-consuming and above all their reproducibility depends on the operator skills. It requires intensive resource involvement. For epidemiological studies these methods are not suitable. The aim of our study was to clinically evaluate the validity and reproducibility of a new automatic device for measurement of pulse wave velocity that can be used in such studies.
Methods:
In 44 subjects including normal healthy control and patients with coronary artery disease, heart brachial, heart ankle, brachial ankle and carotid femoral pulse wave velocities were recorded by using a new oscillometric device. Lead I and II electrocardiogram and pressure curves were simultaneously recorded. Two observers recorded the pulse wave velocity for validation and one observer recorded the velocity on two occasions for reproducibility.
Results And Discussion:
Pulse wave velocity and arterial stiffness index were recorded in 24 control and 20 coronary artery disease patients. All the velocities were significantly high in coronary artery disease patients. There was highly significant correlation between the values noted by the two observers with low standard deviation. The Pearson's correlation coefficient for various velocities ranged from (r = 0.88-0.90) with (p < 0.0001). The reproducibility was also very good as shown by Bland-Altman plot; most of the values were lying within 2 SD. The interperiod measurements of pulse wave velocity were also significantly correlated (r = 0.71-0.98) (P < 0.0001). Carotid-femoral pulse wave velocity was found to correlate significantly with heart brachial, heart ankle, brachial ankle pulse wave velocity and arterial stiffness index values. Reproducibility of our method was good with very low variability in both interobserver and interperiod analysis.
Conclusion:
The new device "PeriScope" based on oscillometric technique has been found to be a simple, non-invasive and reproducible device for the assessment of pulse wave velocity and can be used to determine arterial stiffness in large population based studies.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Pulse
The pulse serves as a clinical indicator...