Using a triple-quadrupole mass spectrometer in accurate mass mode and an ion correlation program to identify
Andrew H Grange1, Witold Winnik, Patrick L Ferguson
1US EPA, ORD, NERL, Environmental Sciences Division, PO Box 93478, Las Vegas, NV 89193-3478, USA. grange.andrew@epa.gov
Rapid Communications in Mass Spectrometry : RCM
|August 27, 2005
Summary
Accurate atomic mass and relative isotopic abundance (RIA) measurements effectively identify ion compositions. This method overcomes interferences, enabling unique compound identification without spectral libraries.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Chemical Analysis
Background:
- Atomic masses and isotopic abundances are crucial for distinguishing ion compositions.
- Accurate mass measurements and relative isotopic abundance (RIA) data significantly reduce possible ion formulas.
- Inherent interferences in mass spectra can distort RIA measurements.
Purpose of the Study:
- To evaluate the efficacy of combining accurate mass and RIA measurements for unique ion composition determination.
- To address and overcome inherent interferences affecting RIA measurements in mass spectrometry.
- To demonstrate a method for compound identification in the absence of commercial spectral libraries.
Main Methods:
- Utilized a triple-quadrupole mass spectrometer with accurate mass capability and atmospheric pressure chemical ionization (APCI).
- Measured exact masses and RIAs of precursor and fragment ions for nine simulated unknown chemicals.
- Employed an ion correlation program to reconcile elemental compositions between precursor ions, fragment ions, and neutral losses.
Main Results:
- Accurate mass (6-10 mDa) and RIA (10%) measurements were sufficient for unique precursor ion composition determination.
- The ion correlation program effectively managed inherent interferences, allowing for accurate RIA analysis.
- Unique compositions were determined for all precursor ions and most fragment ions, compensating for the lack of spectral libraries.
Conclusions:
- Combining accurate mass and RIA measurements provides a robust method for ion composition determination.
- The developed approach offers a viable alternative for compound identification when spectral libraries are unavailable.
- This technique enhances the reliability of mass spectrometry in analyzing complex chemical samples.