Related Experiment Video
Updated: Aug 2, 2026

Non-restraining EEG Radiotelemetry: Epidural and Deep Intracerebral Stereotaxic EEG Electrode Placement
Published on: June 25, 2016
An automatic warning system for epileptic seizures recorded on intracerebral EEGs
1Montreal Neurological Institute, McGill University, Montréal, Qué., Canada H3A 2B4.
Objective:
A new clinical seizure waning system for intracerebral EEG is proposed. It is aimed at a better performance than existing systems and at user tuneability.
Methods:
The system employs data filtering in multiple bands, spectral feature extraction, Bayes' theorem, and spatio-temporal analysis. The a priori information in Bayes' theorem was provided by 407 h of EEG from 19 patients having 152 seizures.
Results:
The testing data (19 patients, 389 h, 100 seizures, independent of the training data) yielded a sensitivity of 89.4%, a false detection rate of 0.22/h, and median delay time of 17.1 s when tuning was used, and 86%, 0.47/h and 16.2 s without tuning. Missed seizures were of short duration or had subtle seizure activity. False detections were caused by technical artefacts, non-epileptic large amplitude rhythmic bursts or very low amplitude activity. It was established that performance could easily be tuned. Results were also compared to the clinical system of .
Conclusions:
The system offers a performance that is acceptable for clinical use. User tuneability allows for reduction in false detection with minimal loss to sensitivity.
Significance:
Epilepsy monitoring generates large amounts of recordings and requires intense observation. Automatic seizure detection and warning systems reduce review time and facilitate observation. We propose a method with high sensitivity and few false alarms.