Related Experiment Video
Updated: Aug 6, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Hydrophilic-coated catheters for intermittent catheterisation reduce urethral micro trauma: a prospective,
J Stensballe1, D Looms, P N Nielsen
1Department of Anaesthesia, Centre of Head and Orthopaedics, Copenhagen University Hospital, Denmark. jakob.stensballe@rh.hosp.dk
Objective:
To compare two hydrophilic-coated (SpeediCath and LoFric and one uncoated gel-lubricated catheter (InCare Advance Plus) concerning withdrawal friction force and urethral micro trauma.
Methods:
49 healthy male volunteers participated in this prospective, randomised, blinded, crossover study of three different bladder catheters. The withdrawal friction force was measured, and urine analysis of blood, nitrite and leucocytes, microbiological analysis of urine cultures and subjective evaluation of the catheters were performed.
Results:
40 participants completed the study and were included in the analysis. SpeediCath exerted a significantly lower mean withdrawal friction force and work than the gel-lubricated uncoated catheter, whereas LoFric exerted a significantly higher mean friction force than both of the other catheters. The hydrophilic catheters caused less microscopic haematuria and less pain than the gel-lubricated uncoated catheter. Furthermore, 93% of the participants preferred the hydrophilic catheters.
Conclusion:
Hydrophilic-coated catheters perform better than uncoated catheters with regard to haematuria and preference. SpeediCath, but not LoFric, exerts less withdrawal friction force than InCare Advance Plus.