Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
[Determination of I- by the ECL of luminol in neutral medium]
Hai-Hong Chu1, Ying-Ying Qi, Yang Xu
1College of Chemistry and Chemical Engineering, Suzhou University, Suzhou 215006, China.
Abstract:
Based on the studies of electrochemiluminescence(ECL) of luminol in neutral medium, the iodide would greatly sensitize the ECL of luminol. The ECL luminous intensity responded linearly to the concentration of iodide within the range of 3.8 x 10(-7) mol x L(-1) to 2.2 x 10(-6) mol x L(-1), and the detection limit is 4.0 x 10(-8) mol x L(-1). The mechanism of enhancement effect of iodide on luminol's ECL is also discussed. The experimental results showed that it was a free radical procedure.