[Effect of the temperature of solution on the emission intensity of ICP]
Jing Ha1, Jin-Zhong Chen, Yan-Hong Wei
1College of Physics Science and Technology, Hebei University, Baoding 071002, China.
Abstract:
The effect of solution temperature (20, 40 and 60 degrees C) on the emission intensity of ICP has been studied. The results show that the spectrum intensity of the elements Ba, Cu and Zn increases obviously with increasing the solution temperature. In addition, at different solution temperatures, the rule of the variation of the intensity with both observation height and nebulizer gas pressure has been obtained, which indicates that the best observation location is lowered when increasing the temperature, but the best nebulizer gas pressure is augmented.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Emission Spectroscopy: Lab
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.