Related Experiment Video
Updated: Aug 16, 2026

Transauricular Vagus Nerve Stimulation and Electroencephalographic Assessment in Disorders of Consciousness
Published on: July 11, 2025
[Effects of desflurane on auditory evoked potential index and quantitative electroencephalogram]
Yu-Guang Xu1, Qu-Lian Guo, Yan-Qing Wang
1Department of Anesthesiology, Xiangya Hospital, Central South University, Changsha 410008, China.
Objective:
To observe the effects of desflurane on A-line ARX Index (AAI), QEEG, MAP and HR, and to investigate the feasibility of AAI, QEEG, MAP and HR in monitoring the depth of anesthesia with desflurane.
Methods:
Thirty patients classified as ASA physical status I approximately II were scheduled for elective lower abdominal surgery. Anesthesia was induced with propofol and vecuronum. After the tracheal intubation, the lungs were ventilated with desflurane in oxygen. End-tidal desflurane concentrations were maintained at 0.8, 1.0 and 1.3 MAC for 20 minutes, respectively. The parameters of record included AAI, BIS (bispectral index), SEF (95% spectral edge freqency), MF (midian freqency), MAP and HR.
Results:
AAI, BIS, SEF and MF significantly decreased in a linear manner (r = 0.830, 0.930, 0.803, and 0.885, respectively, P < 0.01) with increasing end-tidal concentration of desflurane range of 0.8 approximately 1.3 MAC. MAP and HR did not change much. HR increased significantly at concentration 1.3 MAC than that of 1.0 MAC.
Conclusion:
Within end-tidal desflurane concentration range 0.8 approximately 1.3 MAC, both AAI and parameters derived from EEG (such as BIS, SEF, MF) can serve as parameters in monitoring the depth of anesthsia with desflurane, while MAP and HR can not.