Inspiratory oscillatory flow with a portable ventilator: a bench study
Guenther E Frank1, Helmut Trimmel, Robert D Fitzgerald
1Department of Anaesthesiology and Intensive Care, General Hospital Barmherzige Brüder Eisenstadt, Austria. guenther.frank@bbeisen.at
Introduction:
We observed an oscillatory flow while ventilating critically ill patients with the Dräger Oxylog 3000 transport ventilator during interhospital transfer. The phenomenon occurred in paediatric patients or in adult patients with severe airway obstruction ventilated in the pressure-regulated or pressure-controlled mode. As this had not been described previously, we conducted a bench study to investigate the phenomenon.
Methods:
An Oxylog 3000 intensive care unit ventilator and a Dräger Medical Evita-4 NeoFlow intensive care unit ventilator were connected to a Dräger Medical LS800 lung simulator. Data were registered by a Datex-S5 Monitor with a D-fend flow and pressure sensor, and were analysed with a laptop using S5-Collect software. Clinical conditions were simulated using various ventilatory modes, using various ventilator settings, using different filters and endotracheal tubes, and by changing the resistance and compliance. Data were recorded for 258 combinations of patient factors and respirator settings to detect thresholds for the occurrence of the phenomenon and methods to overcome it.
Results:
Under conditions with high resistance in pressure-regulated ventilation with the Oxylog 3000, an oscillatory flow during inspiration produced rapid changes of the airway pressure. The phenomenon resulted in a jerky inspiration with high peak airway pressures, higher than those set on the ventilator. Reducing the inspiratory flow velocity was effective to terminate the phenomenon, but resulted in reduced tidal volumes.
Conclusion:
Oscillatory flow with potentially harmful effects may occur during ventilation with the Dräger Oxylog 3000, especially in conditions with high resistance such as small airways in children (endotracheal tube internal diameter <6 mm) or severe obstructive lung diseases or airway diseases in adult patients.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...