Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution
James Pittman1, Shahar Bar-Yosef, John SumPing
1Duke University Medical Center and Durham Veterans Affairs Medical Center, Durham, NC, USA.
Objective:
Pulse contour analysis can be used to provide beat-to-beat cardiac output (CO) measurement. The current study sought to evaluate this technique by comparing its results with lithium dilution CO (LiCO) measurements.
Design:
Prospective, observational study.
Setting:
Surgical intensive care unit.
Patients:
Twenty-two patients after cardiac or major noncardiac surgery.
Measurements:
After initial calibration of the pulse contour CO (PCO) method, CO was measured by PCO and by LiCO methods at 4, 8, 16, and 24 hrs. Recalibration of PCO was performed every 8 hrs. The systemic vascular resistance and dynamic response characteristics of the arterial catheter-transducer system were measured at each time point to determine whether these influenced the agreement between PCO and LiCO methods.
Main Results:
There was an excellent correlation between methods (r = .94). Bias was small (-0.005 L/min), and clinically acceptable limits of agreement were demonstrated between techniques. Although many catheter-transducer systems had poor dynamic response characteristics, this did not influence the level of agreement between the two techniques. An increase in systemic vascular resistance between two time points did tend to cause overestimation of LiCO by the PCO.
Conclusions:
PCO measurement compared well with the lithium dilution method and can be considered an accurate technique for measuring beat-to-beat CO with limited risk to the patient.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Controlled-Current Coulometry: Overview
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...