Related Experiment Video
Updated: Aug 16, 2026

Post-Myocardial Infarction Heart Failure in Closed-chest Coronary Occlusion/Reperfusion Model in Göttingen Minipigs and Landrace Pigs
Published on: April 17, 2021
Vascular endoscopic and macroscopic observations after crush stenting of coronary artery bifurcations in pigs
Yoshinobu Murasato1, Hiroshi Suzuka, Fumihiko Kamezaki
1Division of Cardiovascular Medicine, Chikuho Social Insurance Hospital, Nogata, Japan. y.murasato@chikuhou-hp.jp
Abstract:
The crush stent technique has recently been proposed to limit the development of restenosis between drug-eluting stents implanted at coronary artery bifurcations. We studied the stent expansion, apposition to the vessel, and aspect of the overlapping stents after in vivo crush stent implantation. Crush stent implantation was performed at coronary bifurcations in anesthetized swines. The treated sites were examined using intravascular ultrasound and a vascular endoscope. The stents removed from the vessel were analyzed macroscopically. After final kissing balloon inflation, an adequate apposition of the stent to the vessel wall was confirmed by vascular endoscopy and visual inspection. However, the side-branch stent was narrowed at the site of stent overlap, and the overlapping stents in the main branch created a metal mass, which could promote the development of thrombosis. The technique of crush stent implantation with additional kissing balloon inflation is feasible and promising. However, it may be limited by thrombosis and restenosis at the carina because of stent overlapping and potential incomplete apposition. Additional studies are needed to confirm the safety and long-term clinical results of this technique.
