A High-throughput Method for Measurement of Glomerular Filtration Rate in Conscious Mice
Published on: May 10, 2013
Variation in the serum creatinine assay calibration: a practical application to glomerular filtration rate estimation
Kalyani Murthy1, Lesley A Stevens, Paul C Stark
1Division of Nephrology, Tufts-New England Medical Center, Boston, Massachusetts, USA. kalyani.murthy@lahey.org
Background:
Variation among clinical laboratories in calibration of serum creatinine assays is a source of error in glomerular filtration rate (GFR) estimation equations. We evaluated impact of this variation on GFR estimates.
Methods:
Errors in GFR estimates were computed based on the range of calibration differences from the 1994 College of American Pathologists (CAP) survey using the Modification of Diet in Renal Disease (MDRD) Study GFR equation.
Results:
Mean (95% CI) calibration difference observed in the CAP survey was +0.14 mg/dL (+12.4 micromol/L) [-0.09, +0.37 mg/dL (-7.96, +32.71 micromol/L)]. Errors in GFR estimates using uncalibrated serum creatinine values were lower in individuals with lower estimated GFR. For GFR of 60 mL/min/1.73 m(2), the mean calibration difference in the CAP survey was associated with errors in GFR estimation between -5.5 to -8.1 mL/min/1.73 m(2) (-9.1 to -13.5%) depending on race and sex. The 95% confidence interval for the calibration difference was associated with a maximal range of error in GFR estimates from +4.6 to -18.1 mL/min/1.73 m(2) (+7.6 to -30.2%). Errors of this magnitude at an estimated GFR of 60 mL/min/1.73 m(2) are not likely to be of clinical significance. However, errors at higher levels of estimated GFR would be greater, making GFR estimates in this range unreliable.
Conclusion:
Recalibration of serum creatinine assays to the MDRD Study clinical laboratory would improve accuracy of GFR estimation using the MDRD Study equation, but is not practical for all clinical laboratories. As an interim solution, clinical laboratories could report GFR estimates <60 mL/min/1.73 m(2) without recalibration with an acceptable accuracy.
Related Concept Videos
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Serum Studies: Renal Function Tests
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...