Effect of 3T MRI on the function of shunt valves--evaluation of Paedi GAV, Dual Switch and proGAV
D Lindner1, C Preul, C Trantakis
1Department of Neurosurgery, University of Leipzig, Liebigstrasse 20, 04103 Leipzig, Germany. dlind@medizin.uni-leipzig.de
Abstract:
The MR-compatibility of medical implants and devices becomes more and more important with the increasing number of high-field MR-scanners employed. Until the end of 2004, about twenty 3T MR in Germany will be in clinical practice. Patients with hydrocephalus need frequent follow-up MR-examinations to assure correct functioning of a shunt. We tested three types of gravitational valves: the Paedi GAV, the Dual Switch and as a new programmable valve the proGAV (Miethke Company, Berlin), that have not been evaluated at 3T, yet. In sum, there is strong evidence for maintenance of function of these valves after exposure to 3T. This also implies the programmable valve, as long as the brake mechanism is properly adjusted during MR-examination.
Related Concept Videos
Shunt Admittances
To model this effect, the method of images is employed. This method involves replacing the Earth with an image conductor that mirrors the original...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Mitral Regurgitation II: Clinical Features and Diagnostic Tests
Mitral Regurgitation III: Medical Management
Mitral Stenosis II: Clinical features and Diagnostic Tests
Mitral Stenosis III: Medical Management