Experimental evaluation of a portable indirect flat-panel detector for the pediatric chest: comparison with storage
Ulrike Rapp-Bernhardt1, Thomas M Bernhardt, Horst Lenzen
1Department of Clinical Radiology, Section of Pediatric Radiology, University of Muenster, 48129 Muenster, Germany.
Purpose:
To compare the exposure dose requirements and performance of a portable indirect flat-panel detector for pediatric use in the depiction of catheters, simulated pulmonary nodules, and simulated interstitial lung disease with those of storage phosphor radiography.
Materials And Methods:
Catheters and simulated nodules and subtle interstitial lung disease (miliary, reticular, linear, and ground-glass patterns) were superimposed over an anthropomorphic chest phantom. Images were obtained with different exposures corresponding to simulated speeds of 400 and 800 with a portable flat-panel detector and printed on hard copies. These images were compared with those from storage phosphor radiography at a simulated speed of 400, which is typically used in pediatric radiology. Four independent readers recorded 7200 observations per pattern (for a total of 600 statistically independent observations), and these observations were subjected to receiver operating characteristic (ROC) analysis. Differences were considered significant at a P value of .05.
Results:
Catheters over obscured chest areas, nodules 10 mm or smaller and larger than 10 mm over lucent lung, nodules 10 mm or smaller over obscured chest areas, and miliary and linear patterns over lucent lung showed higher areas under the ROC curve (A(z)) with the flat-panel detector at 400 and 800 digital speed compared with storage phosphor radiography. A(z) values for reticular and ground-glass patterns with the flat-panel detector were equal to or less than those with storage phosphor radiography. These differences, however, were not statistically significant.
Conclusion:
In the detection of catheters, nodules, and almost all interstitial lung disease, A(z) values were higher with the portable flat-panel detector than with storage phosphor radiography at equivalent and reduced speeds. These results suggest that the portable flat-panel detector could be used with reduced exposure dose in pediatric patients.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...