Background:
Hemodynamic improvement from biventricular pacing is well documented; however, its electrophysiologic effects have not been systematically studied. Sporadic case reports suggest a proarrhythmic effect of biventricular pacing resulting primarily in polymorphic ventricular tachycardia/ventricular fibrillation (VT/VF).
Objectives:
The purpose of this study was to report a series of patients in whom implantation of a biventricular system resulted in VT/VF storm with predominance of monomorphic VT.
Methods:
In a retrospective analysis of all biventricular implants over a 4-year period at a single medical center, we identified 5 of 145 patients (3.4%) who developed VT/VF after they were upgraded to a biventricular system. All patients were male, age 71 +/- 8 years, with ejection fraction of 0.25 +/- 0.1. Four of five patients had ischemic cardiomyopathy.
Results:
All patients developed incessant VT/VF within 1 week of implantation. Monomorphic VT of single morphology was noted in 3 of 5 patients, monomorphic VT of multiple morphologies in 1, and polymorphic VT/VF in 1. VT was managed by temporary discontinuation of biventricular pacing in all patients, amiodarone in 3 of 5, sotalol in 1, and beta-blocker in 1. During 11 +/- 7 months of follow-up, 4 of 5 patients remain alive and are arrhythmia-free.
Conclusion:
Biventricular pacing may result in precipitation of VT/VF storm in a minority of patients with prior history of VT/VF. This may be the first case series reporting both monomorphic and polymorphic VT after an upgrade to a system with biventricular pacing capabilities. The arrhythmias can be managed by conventional therapy and may require temporary discontinuation of left ventricular pacing. This observation is relevant to patients receiving a biventricular pacemaker without an implantable cardioverter-defibrillator backup.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrophysiology of Normal Cardiac Rhythm
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...