Best use of the recommended IFCC reference method, material and values in HbA1C analyses
I M Penttilä1, T Halonen, K Punnonen
1Department of Clinical Chemistry, University Hospital, Kuopio, Finland.
Scandinavian Journal of Clinical and Laboratory Investigation
|September 24, 2005
Summary
Finnish HbA1C surveys show improved accuracy over 10 years. New IFCC standards offer lower values than DCCT, suggesting calibration adjustments for clinical use.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Diabetes Diagnostics
Background:
- Continuous improvement in Finnish HbA1C surveys over 10 years, reducing coefficient of variation.
- Most HbA1C methods trace to the Diabetes Control and Complication Trial (DCCT) method, a now-discontinued HPLC technique.
- The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) developed new reference materials and a method for HbA1C determination.
Purpose of the Study:
- To evaluate the correlation between Labquality Ltd. HbA1C survey results and established reference methods.
- To compare DCCT-aligned values with the new IFCC reference method values.
- To propose a calibration strategy for current methods using IFCC standards.
Main Methods:
- Analysis of Finnish HbA1C survey data from Labquality Ltd. over the past decade.
- Correlation analysis of Labquality Ltd. mean HbA1C values with The European Reference Laboratory for Glycohemoglobin and the IFCC reference method.
- Comparison of IFCC reference values (2.85–3.81%) with current DCCT values (4.0–6.1%).
Main Results:
- Significant improvements in HbA1C measurement consistency observed in Finnish surveys.
- Labquality Ltd. mean HbA1C values showed a near-perfect correlation (r = 0.999) with both European and IFCC reference values.
- IFCC reference values are substantially lower than current DCCT values.
Conclusions:
- The IFCC reference method provides a reliable standard for HbA1C measurement.
- A significant discrepancy exists between IFCC and DCCT values, necessitating careful interpretation.
- Calibration of current methods using IFCC reference materials to DCCT levels is recommended for clinical adoption.