A Method for Systematic Electrochemical and Electrophysiological Evaluation of Neural Recording Electrodes
Published on: March 3, 2014
In vitro comparison of the charge-injection limits of activated iridium oxide (AIROF) and platinum-iridium
Stuart F Cogan1, Philip R Troyk, Julia Ehrlich
1EIC Laboratories Inc, 111 Downey St, Norwood, MA 02062, USA. scogan@eiclabs.com
Abstract:
The charge-injection limits of activated iridium oxide electrodes (AIROF) and PtIr microelectrodes with similar geometric area and shape have been compared in vitro using a stimulation waveform that delivers cathodal current pulses with current-limited control of the electrode bias potential in the interpulse period. Charge-injection limits were compared over a bias range of 0.1-0.7 V (versus Ag/AgCl) and pulse frequencies of 20, 50, and 100 Hz. The AIROF was capable of injecting between 4 and 10 times the charge of the PtIr electrode, with a maximum value of 3.9 mC/cm2 obtained at a 0.7 V bias and 20 Hz frequency.