Chest radiography with a flat-panel detector: image quality with dose reduction after copper filtration
Okka W Hamer1, Claude B Sirlin, Michael Strotzer
1Department of Diagnostic Radiology, University Hospital of Regensburg, Franz-Josef-Strauss-Allee 11, 93042 Regensburg, Germany. o.hamer@gmx.de
Purpose:
To compare image quality and estimated dose for chest radiographs obtained by using a cesium iodide-amorphous silicon flat-panel detector at fixed tube voltage and detector entrance dose with and without additional 0.3-mm copper filtration.
Materials And Methods:
The study was approved by the institutional ethics committee. All prospectively enrolled patients signed the written consent form. Chest radiographs in two projections were acquired at 125-kVp tube voltage and 2.5-microGy detector entrance dose. The experimental group (38 patients) was imaged with 0.3-mm copper filtration; the control group (38 patients) was imaged without copper filtration. An additional 12 patients were imaged with and without copper filtration and served as paired subject-controls. Three readers blinded to group and clinical data independently evaluated the radiographs for image quality on a digital display system. Twelve variables (six for each radiographic projection) were assigned scores on a seven-point ordinal scale. Scores between experimental and control groups were compared: Logistic regression analysis and Mann-Whitney U test were used for unpaired patients; and Wilcoxon and McNemar test, for paired patients. In all, 72 comparisons were determined (36 [12 variables x three readers] for unpaired patients and 36 for paired patients). In a phantom study, radiation burden of experimental protocol was compared with that of control protocol by using Monte Carlo calculations.
Results:
For 70 of 72 comparisons, digital radiographs obtained with copper filtration were of similar image quality as radiographs obtained without copper filtration (P = .123 to P > .99). For two of 72 comparisons, one observer judged the experimental protocol superior to the control protocol (P = .043, P = .046). Patient dose reduction estimated with Monte Carlo calculations was 31%. Use of copper filtration increased exposure times by 48% for posteroanterior views and by 34% for lateral views.
Conclusion:
Subjectively equivalent chest radiographic image quality was found with estimated 30% dose reduction after addition of 0.3-mm copper filtration with flat-panel cesium iodide-amorphous silicon technology.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography