Related Experiment Video
Updated: Aug 15, 2026

The Diffusion of Passive Tracers in Laminar Shear Flow
Published on: May 1, 2018
Study on the impacting factors of transverse diffusion in the micro-channels of T-sensors
Ruijin Wang1, Jianzhong Lin, Zhi-Hua Li
1Department of Mechanics, State Key Laboratory of Fluid Power Transmission and Control, Zhejiang University, Hangzhou 310027, P R. China.
Abstract:
A diffusion-based microfluidic T-sensor with three inlets in which various species are injected was simulated numerically. The results show that the Reynolds number, the dimension, and shape of the micro-channel of the T-sensor, play an important role in the diffusion of flow. The smaller the channel width is, the wider the species diffuse. The inlet angle larger than 45 degrees will not exert any influence on the diffusion. The serpentine channel is a better structure for transverse diffusion and mixing in the micro-channel. These conclusions are helpful for the design and fabrication of micro-fluidic devices as well as the analysis of data collected from such devices.