A commercial whole blood glucose biosensor with a low sensitivity to hematocrit based on an impregnated porous carbon

Nigel J Forrow1, Simon W Bayliff

  • 1MediSense Products, Abbott Diabetes Care, Abbott Laboratories, Range Road, Witney, Oxon OX29 0YL, UK. nigel.forrow@abbott.com

Erythrocytes (red blood cells) are a major source of response variation in biosensor electrodes expected to operate in whole blood. Such a blood-to-plasma difference (hematocrit effect) must be minimized for those sensors directed towards the hospital market where wide variations in hematocrit can be seen. Typically, many current glucose sensors demonstrate a decreasing response to the analyte in the presence of increasing hematocrit levels. A sensor electrode for glucose is described which displays a reduced sensitivity to changes in hematocrit. The working electrode comprises a base porous conducting carbon layer, which is impregnated with a mixture including glucose oxidase and a ferrocene redox mediator. The base carbon layer has a void volume of 50%, an average pore diameter of less than 0.1 microm and a thickness of about 20 microm. The interior void volume of the base carbon layer is filled entirely with a substantial proportion of the impregnating mixture such that very little remains on the exterior. The resulting impregnated porous electrode excludes erythrocytes and is consequently capable of operating acceptably in venous, capillary, arterial and neonatal blood over a wide hematocrit range of 20-70%.

Related Concept Videos

Amperometry: Overview01:10

Amperometry: Overview

Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
Potentiometry: Membrane Electrodes01:15

Potentiometry: Membrane Electrodes

Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at the...
Microbial Biosensors01:17

Microbial Biosensors

Microbial biosensors are analytical devices that utilize living microbes to detect specific substances through measurable signals. These devices consist of two main components: biosensing organisms and signal-transducing elements. Biosensing organisms, such as Escherichia coli or Saccharomyces cerevisiae, are typically housed in multiwell plates connected to transducers, enabling rapid, real-time detection of target analytes.Signal Generation MechanismWhen a target analyte—such as...