Related Experiment Video
Updated: Aug 13, 2026

Multiplexed Fluorescent Immunohistochemical Staining, Imaging, and Analysis in Histological Samples of Lymphoma
Published on: January 9, 2019
Antibody selection in immunohistochemical detection of cyclin D1 in mantle cell lymphoma
Emina Torlakovic1, Søren Nielsen, Mogens Vyberg
1Department of Pathology, Royal University Hospital, College of Medicine, University of Saskatchewan, Saskatoon, Canada.
Abstract:
An assessment in Nordic immunohistochemical Quality Control (NordiQC) revealed that only 23% participant laboratories performed optimal staining for detection of cyclin D1 (CyD1) in mantle cell lymphoma (MCL). False-negative results were secondary to suboptimal protocols. We compared the 5 anti-CyD1 antibodies (monoclonal SP4, P2D11F11, and DCS-6 and polyclonal CP236 and 06-137) used in the Scandinavian laboratories. Evaluated were 31 MCLs, 16 other malignant lymphomas, and 19 samples of normal tissues. Sensitivity was as follows: CP236, 100%; SP4, 95%; P2D11F11, 90%; DCS-6, 84%; and 06-137, 53%. SP4 produced the strongest staining. Correlation of CyD1 with the proliferative index was best with polyclonal antibodies CP236 and 06-137. The use of heat-induced epitope retrieval in an alkaline buffer such as 1/10 mmol/L of Tris-EDTA buffer, pH 9, seemed mandatory. For the optimal detection of CyD1 expression, both SP4 and CP236 antibodies should be available in the laboratory.

