Related Experiment Video
Updated: Aug 15, 2026

Three-Dimensionally Printed Microfluidic Cross-flow System for Ultrafiltration/Nanofiltration Membrane Performance Testing
Published on: February 13, 2016
Middle molecule removal in low-flux polysulfone dialyzers: impact of flows and surface area on whole-body and
Sunny Eloot1, Jean-Yves de Vos, Filip de Vos
1Institute Biomedical Technology, Ghent University, Gent, Belgium. sunny.eloot@navierugent.be
Abstract:
Some studies found that the removal of middle molecules has a long-term effect on mortality and, even more, is enhanced by high-flux dialysis. In order to enhance middle molecule removal in a low-flux dialyzer, the present study aimed at investigating the combined impact of dialyzer flows and membrane surface area. Blood and dialysate flows were varied within the clinical range 300-500 and 500-800 mL/min, respectively, while the ultrafiltration rate was kept constant at 0.1 L/hr. Single-pass tests were performed in vitro in a single Fresenius F6HPS dialyzer (3 tests) and serially (5 tests) and parallel (3 tests) connected dialyzers. The blood substitution fluid consisted of dialysis fluid in which radioactive-labeled vitamin B12 (molecular weight 1355 Da) was dissolved. Dialyzer clearance as well as whole-body clearance was calculated from radioactivity concentrations of samples taken from the inlet and outlet bloodline. Adding a second dialyzer in series or parallel ameliorated the overall dialyzer and whole-body clearance significantly, except for the highest applied blood flows of 500 mL/min. Better solute removal was also obtained with higher dialysate flows, while the use of higher blood flows seemed advantageous only when using a single dialyzer. Analysis of the ultrafiltration profiles in the different configurations illustrated that enhancing the internal filtration rate ameliorates convective transport of middle molecules. Adequate solute removal results from a number of interactions, as there are blood and dialysate flows, membrane surface area, filtration profile and concentration profiles in the blood and dialysate compartment.
More Related Videos
Related Concept Videos
Dialysis
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Peritoneal Dialysis I: Introduction and Procedure
Factors Affecting Renal Clearance: Drug's Physicochemical Properties and Plasma Levels
One important factor is the drug's molecular size. The kidneys readily excrete smaller molecules below 300 Daltons (Da). On the other hand, molecules weighing between 300 and 500 Da are excreted through both urine and bile. Larger molecules above 500 Da tend to be excreted less...
Clearance Models: Physiological Models
The organ's clearance rate depends on the blood flow to the organ and the extraction ratio (E). The extraction ratio describes the organ's proficiency in drug...
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis